Accéder directement au contenu Accéder directement à la navigation
Article dans une revue

Automatic cytometric device using multiple wavelength excitations

Abstract : Precise identification of eosinophils, basophils, and specific subpopulations of blood cells (B lymphocytes) in an unconventional automatic hematology analyzer is demonstrated. Our specific apparatus mixes two excitation radiations by means of an acousto-optics tunable filter to properly control fluorescence emission of phycoerythrin cyanin 5 (PC5) conjugated to antibodies (anti-CD20 or anti-CRTH2) and Thiazole Orange. This way our analyzer combining techniques of hematology analysis and flow cytometry based on multiple fluorescence detection, drastically improves the signal to noise ratio and decreases the spectral overlaps impact coming from multiple fluorescence emissions.
Type de document :
Article dans une revue
Liste complète des métadonnées

https://hal-unilim.archives-ouvertes.fr/hal-00643743
Contributeur : Véronique Maury <>
Soumis le : mardi 22 novembre 2011 - 16:48:07
Dernière modification le : jeudi 11 janvier 2018 - 06:26:10

Identifiants

  • HAL Id : hal-00643743, version 1

Collections

Citation

Nelly Rongeat, S. Ledroit, L. Chauvet, D. Cremien, Alexandra Urankar, et al.. Automatic cytometric device using multiple wavelength excitations. Journal Biomedical optics, 2011, 16 (5), pp.PP. 057006. ⟨hal-00643743⟩